Common Measurement Settings
R&S
®
FSW
372
User Manual 1173.9411.02 ─ 19
Write to <CVL table name> ← Bias Settings
Stores the bias setting in the currently selected "Conversion loss table" for the range
(see
"Managing Conversion Loss Tables"
on page 372). If no conversion loss table is
selected yet, this function is not available ("CVL Table not selected").
Remote command:
Managing Conversion Loss Tables
In this tab you configure and manage conversion loss tables. Conversion loss tables
consist of value pairs that describe the correction values for conversion loss at certain
frequencies. The correction values for frequencies between the reference points are
obtained via interpolation.
The currently selected table for each range is displayed at the top of the dialog box. All
conversion loss tables found in the instrument's
C:\r_s\instr\user\cvl\
direc-
tory are listed in the "Modify Tables" list.
New Table
Opens the "Edit Conversion loss table" dialog box to configure a new conversion loss
table. For details on table configuration see
"Creating and Editing Conversion Loss
Remote command:
Data Input and Output