![Rohde & Schwarz FSW13 User Manual Download Page 867](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847867.webp)
Remote Commands
R&S
®
FSW
867
User Manual 1173.9411.02 ─ 19
Parameters:
<Source>
RF
Radio Frequency ("RF INPUT" connector)
DIQ
Digital IQ data (only available with optional Digital Baseband
Interface R&S
FSW-B17)
For details on I/Q input see the R&S
FSW I/Q Analyzer User
Manual.
AIQ
Analog Baseband signal (only available with optional Analog
Baseband Interface R&S
FSW-B71)
For details on Analog Baseband input see the R&S
FSW I/Q
Analyzer User Manual.
*RST:
RF
Manual operation:
See
INPut:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FSW.
See the R&S
FSW Getting Started manual for details.
Parameters:
<State>
ON | 1
User port is switched to INPut
OFF | 0
User port is switched to OUTPut
*RST:
1
INPut:UPORt[:VALue]?
This command queries the control lines of the user ports.
Example:
INP:UPOR?
Usage:
Query only
11.7.6.2
Using External Mixers
The commands required to work with external mixers in a remote environment are
described here. Note that these commands require the R&S
FSW-B21 option to be
installed and an external mixer to be connected to the front panel of the R&S
FSW. In
MSRA/MSRT mode, external mixers are not supported.
For details on working with external mixers see
chapter 6.2.5.1, "Basics on External
Setting Basic Measurement Parameters