Common Measurement Settings
R&S
®
FSW
428
User Manual 1173.9411.02 ─ 19
(See
Remote command:
Free Run ← Trigger Source ← Trigger Settings
No trigger source is considered. Data acquisition is started manually or automatically
and continues until stopped explicitely.
In the Spectrum application, this is the default setting.
Remote command:
TRIG:SOUR IMM
, see
External Trigger 1/2/3 ← Trigger Source ← Trigger Settings
Data acquisition starts when the TTL signal fed into the specified input connector (on
the front or rear panel) meets or exceeds the specified trigger level.
(See
Note:
The "External Trigger 1" softkey automatically selects the trigger signal from the
TRIGGER INPUT connector on the front panel.
For details see the "Instrument Tour" chapter in the R&S
FSW Getting Started manual.
"External Trigger 1"
Trigger signal from the TRIGGER 1 INPUT connector on the front
panel.
"External Trigger 2"
Trigger signal from the TRIGGER 2 INPUT/OUTPUT connector on
the front panel.
Note: Connector must be configured for "Input" in the "Outputs" con-
figuration (see
"External Trigger 3"
Trigger signal from the TRIGGER 3 INPUT/ OUTPUT connector on
the rear panel.
Note: Connector must be configured for "Input" in the "Outputs" con-
figuration (see
Remote command:
TRIG:SOUR EXT
,
TRIG:SOUR EXT2
TRIG:SOUR EXT3
See
SWE:EGAT:SOUR EXT
for gated triggering, see
Video ← Trigger Source ← Trigger Settings
Defines triggering by the video signal, i.e. the filtered and detected version of the input
signal (the envelope of the IF signal), as displayed on the screen.
Define a trigger level from 0 % to 100 % of the diagram height. The absolute trigger
level is indicated by a horizontal trigger line in the diagram, which you can also move
graphically to change the trigger level.
Trigger and Gate Configuration