Remote Commands
R&S
®
FSW
876
User Manual 1173.9411.02 ─ 19
This command is only available with option B21 (External Mixer) installed.
Usage:
Query only
[SENSe:]CORRection:CVL:CLEAr
This command deletes the selected conversion loss table. Before this command can
be performed, the conversion loss table must be selected (see
This command is only available with option B21 (External Mixer) installed.
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:CLE
Usage:
Event
Manual operation:
See
[SENSe:]CORRection:CVL:COMMent
<Text>
This command defines a comment for the conversion loss table. Before this command
can be performed, the conversion loss table must be selected (see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Text>
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:COMM 'Conversion loss table for
FS_Z60'
Manual operation:
See
[SENSe:]CORRection:CVL:DATA
<Freq>,<Level>
This command defines the reference values of the selected conversion loss tables.
The values are entered as a set of frequency/level pairs. A maximum of 50 frequency/
level pairs may be entered. Before this command can be performed, the conversion
loss table must be selected (see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Freq>
numeric value
The frequencies have to be sent in ascending order.
<Level>
Setting Basic Measurement Parameters