Network and Remote Operation
R&S
®
FSW
635
User Manual 1173.9411.02 ─ 19
Table 10-7: Meaning of the bits used in the STATus:QUEStionable register
Bit No.
Meaning
0 - 1
Unused
2
TIMe
This bit is set if a time error occurs in any of the active channels.
STATus:QUEStionable:TIMe Register
provides more information on the error type.
3
POWer
This bit is set if the measured power level in any of the active channels is questionable.
STATus:QUEStionable:POWer Register
provides more information on the error type.
4
TEMPerature
This bit is set if the temperature is questionable.
5
FREQuency
This bit is set if there is anything wrong with the frequency of the local oscillator or the reference
frequency in any of the active channels.
STATus:QUEStionable:FREQuency Register
provides more information on the error type.
6 - 7
Unused
8
CALibration
This bit is set if the R&S
FSW is unaligned ("UNCAL" display)
9
LIMit (device-specific)
This bit is set if a limit value is violated in any of the active channels in any window.
STATus:QUEStionable:LIMit Register
provides more information on the error type.
10
LMARgin (device-specific)
This bit is set if a margin is violated in any of the active channels in any window.
STATus:QUEStionable:LMARgin Register
provides more information on the error type.
11
SYNC (device-specific)
This bit is set if the R&S
FSW is not synchronized to the signal that is applied.
The R&S
FSW is not synchronized if:
●
it cannot synchronize to midamble during a measurement or premeasurement
●
it cannot find a burst during a measurement or premeasurement
●
the results deviate too much from the expected value during premeasurements
12
ACPLimit (device-specific)
This bit is set if a limit during ACLR measurements is violated in any of the active channels.
STATus:QUEStionable:ACPLimit Register
provides more information on the error type.
13
Unused
14
Digital I/Q (device-specific)
This bit is set if a connection error occurs at the Digital Baseband Interface (R&S
FSW-B17)
For details see the R&S
FSW I/Q Analyzer User Manual.
15
This bit is always 0.