Reference Frequency Input
The R&S
FSW can use the internal reference source or an external reference source
as the frequency standard for all internal oscillators. A 10
MHz crystal oscillator is used
as the internal reference source. In the external reference setting, all internal oscillators
of the R&S
FSW are synchronized to the external reference frequency.
External references are connected to one of the REF INPUT or the SYNC TRIGGER
connectors on the rear panel. For details see the "Getting Started" manual.
Note:
Optionally (R&S
FSW-B4), the more precise OCXO signal can replace the inter-
nal reference source.
The default setting is the internal reference. When an external reference is used, "EXT
REF" is displayed in the status bar.
The following reference inputs are available:
Table 9-1: Available Reference Frequency Input
Source
Frequency Tuning
Range
Loop Band-
width
Description
Internal
10 MHz
-
1-100 Hz
Internal reference signal
or OCXO (option R&S
FSW-B4)
External Reference
10
MHz
10 MHz
+/- 6 ppm
1-100 Hz
External reference from REF INPUT
1..20
MHZ connector;
Fixed external 10
MHZ reference fre-
quency.
Good phase noise performance
External Reference
1..20 MHz
1..20 MHz
in 1
Hz
steps
+/- 0.5 ppm
0.1 Hz (fixed)
Variable external reference frequency
in 0.1
Hz steps from REF INPUT
1..20
MHZ connector;
Good external phase noise suppres-
sion. Small tuning range.
+/- 6 ppm
1-30 Hz
Variable external reference frequency
in 0.1
Hz steps from REF INPUT
1..20
MHZ connector;
Wide tuning range.
External Reference
100
MHz
100 MHz
+/- 6 ppm
1-300 Hz
External reference from REF INPUT
100
MHZ connector
Good phase noise performance
Sync Trigger
100 MHz
+/- 6 ppm
1-300 Hz
External reference from SYNC
TRIGGER INPUT connector
Remote command:
General Instrument Settings