Data Management
R&S
®
FSW
546
User Manual 1173.9411.02 ─ 19
Item
Description
Noise - Calibration data
Results from calibration measurement (R&S
FSW-K30 only)
K40 Results
All current phase noise trace results (R&S
FSW-K40 only)
8.3.2.2
Storage Location and File Name
The data is stored on the internal flash disk or, if selected, on a memory stick or net-
work drive. The operating system, firmware and stored instrument settings are located
on drive C. All other folders and drives can be used to store measurement data.
Saving instrument settings in secure user mode
In secure user mode all data is stored to the SDRAM, and is only available during the
current session. As soon as the power is switched off on the R&S
FSW, the data is
cleared (see
chapter 3.1.7, "Protecting Data Using the Secure User Mode"
on page 37).
Other storage locations cannot be selected in this mode.
The storage location and file name are selected in a file selection dialog box which is
displayed when you perform a storage function.
By default, the name of a settings file consists of a base name followed by an under-
score and three numbers, e.g.
limit_lines_005
. In the example, the base name is
limit_lines
. The base name can contain characters, numbers and underscores.
The file extension
dfl
is added automatically. The default folder for settings files is
C:\r_s\instr\user
.
8.3.2.3
Save and Recall Dialog Boxes
The following dialog boxes are available via softkeys in the "Save/Recall" menu which
is displayed when you select the "Save" or "Open" icon in the toolbar. Both dialog
boxes are very similar and closely related.
Storing and Recalling Instrument Settings and Measurement Data