Remote Commands
R&S
®
FSW
1027
User Manual 1173.9411.02 ─ 19
Parameters:
<ScalingType>
LINear | LOGarithmic
*RST:
LINear
Manual operation:
See
[SENSe:]CORRection:TRANsducer:SELect
<Name>
This command selects a transducer factor.
Parameters:
<Name>
String containing the name of the transducer factor.
If the name does not exist yet, the R&S
FSW creates a trans-
ducer factor by that name.
Example:
CORR:TRAN:SEL 'FACTOR1'
Manual operation:
See
See
See
[SENSe:]CORRection:TRANsducer[:STATe]
<State>
This command turns the selected transducer factor on or off.
Before you can use the command, you have to select a transducer.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
[SENSe:]CORRection:TRANsducer:UNIT
<Unit>
This command selects the unit of the transducer factor.
Before you can use the command, you have to select and turn on a transducer.
Parameters:
<Unit>
string as defined in table below
*RST:
DB
Example:
CORR:TRAN:UNIT 'DBUV'
Manual operation:
See
String
Unit
'DB'
dB
'DBM'
dBm
'DBMV'
dBmV
Configuring the R&S
FSW