General Instrument Setup
R&S
®
FSW
585
User Manual 1173.9411.02 ─ 19
Save Device Footprint
Creates an *.xml file with information on installed hardware, software, image and
FPGA versions. The *.xml file is stored under
C:\R_S\Instr\devicedata\xml\
on
the instrument. It is also included in the
service.zip
file.
9.3.5.2
Selftest Settings and Results
If the R&S
FSW fails you can perform a self test of the instrument to identify any defec-
tive modules.
The selftest settings and results are available in the "Selftest" tab of the "Service" dia-
log box.
Once the self test is started, all modules are checked consecutively and the test result
is displayed. You can abort a running test.
In case of failure a short description of the failed test, the defective module, the associ-
ated value range and the corresponding test results are indicated.
A running Sequencer process is aborted when you start a self-alignment.
Remote command:
General Instrument Settings