Remote Commands
R&S
®
FSW
1024
User Manual 1173.9411.02 ─ 19
Manual operation:
See
DIAGnostic:SERVice:INPut:AIQ[:TYPE]
<SignalType>
This command defines the type of calibration signal to be used for Analog Baseband.
This command is only available if the R&S
FSW-B71 option is installed.
Parameters:
<SignalType>
AC
1.5625 MHz square wave AC signal
DC
DC signal
DCZ
no signal
*RST:
AC
Example:
DIAG:SERV:INP:AIQ:TYPE DCZ
Manual operation:
See
DIAGnostic:SERVice:INPut[:SELect]
<Signal>
This command activates or deactivates the use of an internal calibration signal as input
for the R&S
FSW.
Parameters:
<Signal>
CALibration
Uses the calibration signal as RF input.
MCALibration
Uses the calibration signal for the microwave range as RF input.
RF
Uses the signal from the RF input.
AIQ
Uses the Analog Baseband calibration signal as input to the
optional Analog Baseband interface. This signal is only available
if the R&S
FSW-B71 option is installed.
*RST:
RF
Example:
DIAG:SERV:INP CAL
Uses the calibration signal as RF input.
Manual operation:
See
See
See
See
DIAGnostic:SERVice:STESt:RESult?
This command queries the self test results.