Getting Started
R&S
®
FSW
59
User Manual 1173.9411.02 ─ 19
.........................................................................................68
....................................................................................73
3.3.1 Measuring a Basic Signal
We will start out by measuring a simple sinus wave, using the internal calibration signal
as the input.
To display the internal 64 MHz calibration signal
1. Press the PRESET key to start out in a defined instrument configuration.
2. Press the SETUP key on the front panel.
3. Tap the "Service" softkey.
4. Tap the "Calibration Signal" tab.
5. Tap the "Calibration Frequency RF" option. Leave the frequency at the default
64
MHz, with a narrowband spectrum.
The calibration signal is now sent to the RF input of the R&S
FSW. By default, a
continuous frequency sweep is performed, so that the spectrum of the calibration
signal is now displayed in the standard level versus frequency diagram.
Fig. 3-4: Calibration signal as RF input
Trying Out the Instrument