Getting Started
R&S
®
FSW
68
User Manual 1173.9411.02 ─ 19
3.
Tap the "SmartGrid" icon in the toolbar to activate SmartGrid mode.
4. Drag the "Marker Peak List" icon from the evaluation bar to the lower half of the
display to add a new window for the peak list.
5. Close the SmartGrid mode.
6. To obtain a more conclusive peak list that does not contain noise peaks, for exam-
ple, define a threshold that is higher than the noise floor:
a) Press the MKR key on the front panel.
b) Tap the "Marker Config" softkey in the "Marker" menu.
c) Tap the "Search" tab in the "Marker" dialog box.
d) In the "Threshold" field, enter
-68
dBm
.
e) Tap the "State" box for "Threshold" to activate its use.
Only peaks that are larger than -68
dBm will be included in the peak list.
The marker peak list displays the determined peaks that are above the defined
threshold.
Fig. 3-14: Marker Peak List
3.3.7 Zooming into the Display
In order to analyze the areas around the peak levels in more detail, we will zoom into
the top 3 peaks.
1.
Tap the "Multiple Zoom" icon in the toolbar.
Trying Out the Instrument