![Rohde & Schwarz FSW13 User Manual Download Page 771](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847771.webp)
Remote Commands
R&S
®
FSW
771
User Manual 1173.9411.02 ─ 19
The command requires option R&S
FSW-B24.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Gain>
15 dB | 30 dB
The availability of preamplification levels depends on the
R&S
FSW model.
•
R&S
FSW8/13: 15dB and 30 dB
•
R&S
FSW26 or higher: 30 dB
All other values are rounded to the nearest of these two.
*RST:
OFF
[SENSe:]LIST:RANGe<range>:LIMit:STARt
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Level>
Absolute limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
Manual operation:
See
[SENSe:]LIST:RANGe:LIMit:STATe
This command turns the limit check for all spurious emission measurement ranges on
and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
[SENSe:]LIST:RANGe<range>:LIMit:STOP
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Configuring and Performing Measurements