CALCulate:LIMit<k>:LOWer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Parameters:
<LimitLinePoints>
Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with
FSW either adds
missing values or ignores surplus values.
The unit depends on
on page 982.
*RST:
Limit line state is OFF
Usage:
SCPI confirmed
Manual operation:
See
CALCulate:LIMit<k>:LOWer:MARGin
<Margin>
This command defines an area around a lower limit line where limit check violations
are still tolerated.
Parameters:
<Margin>
numeric value
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate:LIMit<k>:LOWer:MODE
<Mode>
This command selects the vertical limit line scaling.
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
Manual operation:
See
CALCulate:LIMit<k>:LOWer:OFFSet
<Offset>
This command defines an offset for a complete lower limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Analyzing Measurements (Basics)