Network and Remote Operation
R&S
®
FSW
651
User Manual 1173.9411.02 ─ 19
.............................................................................................. 651
GPIB Address
Defines the GPIB address. Values from 0 to 30 are allowed. The default address is 20.
Remote command:
Identification String
Defines the identification string for the device which is provided as a response to the
*IDN?
query. Maximum 36 characters are allowed.
Remote command:
Reset to Factory String
Restores the default identification string. Each instrument has a unique ID according to
the following syntax:
Rohde&Schwarz,FSW,<Unique number>
Remote command:
Remote Display Update
Defines whether the instrument display is updated or switched off when changing from
manual operation to remote control.
Remote command:
GPIB Terminator
Changes the GPIB receive terminator.
According to the standard, the terminator in ASCII is <LF> and/or <EOI>. For binary
data transfers (e.g. trace data) from the control computer to the instrument, the binary
code used for <LF> might be included in the binary data block, and therefore should
not be interpreted as a terminator in this particular case. This can be avoided by
changing the receive terminator to EOI.
Remote command:
*IDN Format
Defines the response format to the remote command
*IDN?
(see
on page 679). This function is intended for re-use of existing control programs
together with the R&S
FSW.
"Leg"
Legacy format, compatible to the R&S FSP/FSU/FSQ family