Remote Commands
R&S
®
FSW
1036
User Manual 1173.9411.02 ─ 19
Example:
SYST:COMM:GPIB:ADDR 18
Usage:
SCPI confirmed
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
<Terminator>
This command selects the GPIB receive terminator.
According to the standard the terminator in ASCII is <LF> and/or <EOI>. For binary
data transfers (e.g. trace data) from the control computer to the instrument, the binary
code (0AH) used for <LF> might be included in the binary data block, and therefore
should not be interpreted as a terminator in this particular case. This can be avoided by
changing the receive terminator to EOI.
Output of binary data from the instrument to the control computer does not require
such a terminator change.
Parameters:
<Terminator>
LFEOI | EOI
*RST:
(no influence on this parameter, factory default
LFEOI)
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation:
See
SYSTem:DISPlay:UPDate
<State>
This command turns the display during remote operation on and off.
If on, the R&S
FSW updates the diagrams, traces and display fields only.
The best performance is obtained if the display is off during remote control operation.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
SYST:DISP:UPD ON
Manual operation:
See
SYSTem:IDENtify:FACTory
This command resets the query to
to its default value.
Usage:
Event
Manual operation:
See
Configuring the R&S
FSW