Remote Commands
R&S
®
FSW
853
User Manual 1173.9411.02 ─ 19
Example:
TRIG:LEV 2V
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event. Note that any RF attenuation or preamplification is
considered when the trigger level is analyzed. If defined, a reference level offset is also
considered.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-10 dBm
Example:
TRIG:LEV:IFP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IQPower
<TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event. Note that any RF attenuation or preamplification is considered when the trigger
level is analyzed.
Parameters:
<TriggerLevel>
Range:
-130 dBm to 30 dBm
*RST:
-20 dBm
Example:
TRIG:LEV:IQP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:LEVel:RFPower
<TriggerLevel>
This command defines the power level the RF input must exceed to cause a trigger
event. Note that any RF attenuation or preamplification is considered when the trigger
level is analyzed. If defined, a reference level offset is also considered.
The input signal must be between 500
MHz and 8
GHz.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-20 dBm
Example:
TRIG:LEV:RFP -30dBm
Manual operation:
See
Setting Basic Measurement Parameters