Measurements
R&S
®
FSW
246
User Manual 1173.9411.02 ─ 19
For details on transducers see
chapter 9.2, "Basics on Transducer Factors"
Remote command:
Limit Check
Activates or deactivates the limit check for all ranges.
For details on limit checks see
chapter 5.7.3.2, "Limit Lines in Spurious Measure-
"Absolute"
Signal is checked against absolute limit values
"None"
No limit check is performed.
Remote command:
Abs Limit Start/Stop
Sets an absolute limit value at the start or stop frequency of the range [dBm].
Remote command:
Insert before/after Range
Inserts a new range to the left of the currently focused range (before) or to the right
(after). The range numbers of the currently focused range and all higher ranges are
increased accordingly. The maximum number of ranges is 30.
Delete Range
Deletes the currently focused range. The range numbers are updated accordingly.
5.7.4.2
Adjusting the X-Axis to the Range Definitions
The frequency axis of the measurement diagram can be adjusted automatically so that
the span of all sweep list ranges corresponds to the displayed span. Thus, the x-axis
range is set from the start frequency of the first sweep range to the stop frequency of
the last sweep range.
5.7.4.3
List Evaluation
In the "List Evaluation" dialog box, which is displayed when you select the "Evalua-
tions" button in the "Overview" or the "List Evaluation" softkey in the "Spurious Emis-
sions" menu, you configure the contents and display of the result list.
Spurious Emissions Measurement