DISPlay[:WINDow<n>]:TRACe<t>:MODE
<Mode>
This command selects the trace mode.
In case of max hold, min hold or average trace mode, you can set the number of single
measurements with
. Note that synchronization to the end of
the measurement is possible only in single sweep mode.
Parameters:
<Mode>
WRITe
Overwrite mode: the trace is overwritten by each sweep. This is
the default setting.
AVERage
The average is formed over several sweeps. The "Sweep/Aver-
age Count" determines the number of averaging procedures.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S
FSW saves the sweep result in the trace mem-
ory only if the new value is greater than the previous one.
MINHold
The minimum value is determined from several measurements
and displayed. The R&S
FSW saves the sweep result in the
trace memory only if the new value is lower than the previous
one.
VIEW
The current contents of the trace memory are frozen and dis-
played.
BLANk
Hides the selected trace.
*RST:
Trace 1: WRITe, Trace 2-6: BLANk
Example:
INIT:CONT OFF
Switching to single sweep mode.
SWE:COUN 16
Sets the number of measurements to 16.
DISP:TRAC3:MODE WRIT
Selects clear/write mode for trace 3.
INIT;*WAI
Starts the measurement and waits for the end of the measure-
ment.
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:MODE:HCONtinuous
<State>
This command turns an automatic reset of a trace on and off after a parameter has
changed.
The reset works for trace modes min hold, max hold and average.
Analyzing Measurements (Basics)