Remote Commands
R&S
®
FSW
878
User Manual 1173.9411.02 ─ 19
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<PortType>
2 | 3
*RST:
2
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:PORT 3
Manual operation:
See
[SENSe:]CORRection:CVL:SELect
<FileName>
This command selects the conversion loss table with the specified file name. If
<file_name> is not available, a new conversion loss table is created.
This command is only available with option B21 (External Mixer) installed.
Parameters:
<FileName>
'<File name>'
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Manual operation:
See
See
See
[SENSe:]CORRection:CVL:SNUMber
<SerialNo>
This command defines the serial number of the mixer for which the conversion loss
table is to be used. This setting is checked against the current mixer setting before the
table can be assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<SerialNo>
Serial number with a maximum of 16 characters
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:MIX '123.4567'
Manual operation:
See
Programming Example: Working with an External Mixer
This example demonstrates how to work with an external mixer in a remote environ-
ment. It is performed in the Spectrum application in the default layout configuration.
Setting Basic Measurement Parameters