Common Measurement Settings
R&S
®
FSW
387
User Manual 1173.9411.02 ─ 19
Thus, for logarithmic x-axis scaling, the number of sweep points must be sufficiently
high in order to distinguish high frequencies precisely. The resolution bandwidth should
cover at least one sweep point (that means: the distance between two sweep points
should not exceed the RBW). If this condition is not met, signals or interferences could
be missed, especially narrowband interferences.
6.3.2 Frequency and Span Settings
Frequency and span settings can be configured via the "Frequency" dialog box. Signal
tracking is configured in the "Signal Tracking" tab of this dialog box. For details see
chapter 6.3.3, "How To Define the Frequency Range"
Frequency and Span Configuration