Remote Commands
R&S
®
FSW
1023
User Manual 1173.9411.02 ─ 19
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage:
Query only
Manual operation:
See
DIAGnostic:SERVice:INPut:MC:DISTance
<Bandwidth>
This command selects the distance of the peaks of the microwave calibration signal for
calibration of the YIG filter.
This command is only available for instrument models R&S
FSW13/26.
Parameters:
<Bandwidth>
SMALl
Small offset of combline frequencies.
WIDE
Wide offset of combline frequencies.
Manual operation:
See
DIAGnostic:SERVice:INPut:PULSed:CFRequency
<Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
.
Parameters:
<Frequency>
Possible frequencies of the calibration signal are fix.
If you define a frequency that is not available, the R&S
FSW
uses the next available frequency. Example: a frequency of
20
MHz is rounded up to the next available frequency
(21.33
MHz).
*RST:
64 MHz
Default unit: Hz
Manual operation:
See
DIAGnostic:SERVice:INPut:RF:SPECtrum
<Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth>
NARRowband
Narrowband signal for power calibration of the frontend.
BROadband
Broadband signal for calibration of the IF filter.
Configuring the R&S
FSW