![Rohde & Schwarz FSW13 User Manual Download Page 772](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847772.webp)
Remote Commands
R&S
®
FSW
772
User Manual 1173.9411.02 ─ 19
Parameters:
<Level>
Absolute limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
Manual operation:
See
[SENSe:]LIST:RANGe<range>:POINts[:VALue]
<Points>
This command defines the number of sweep points in a spurious emission measure-
ment range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Points>
For more information on sweep points see
Much Data is Measured: Sweep Points and Sweep Count"
on page 407.
*RST:
1001
Manual operation:
See
[SENSe:]LIST:RANGe<range>:RLEVel
<RefLevel>
This command defines the reference level for a spurious emission measurement
range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<RefLevel>
Reference level.
Refer to the data sheet for the reference level range.
*RST:
0 dBm
Manual operation:
See
[SENSe:]LIST:RANGe<range>:SWEep:TIME
<SweepTime>
This command defines the sweep time for a spurious emission measurement range.
Suffix:
<range>
.
1..30
Selects the measurement range.
Configuring and Performing Measurements