![Rohde & Schwarz FSW13 User Manual Download Page 46](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847046.webp)
Getting Started
R&S
®
FSW
46
User Manual 1173.9411.02 ─ 19
using the VOLUME control to the right of the female connector. If a plug is connected,
the internal loudspeaker is automatically switched off.
To use the PHONES connector, in the "Output" dialog box (INPUT/OUTPUT key),
select "Video Output" = "Video". The output voltage (volume) is 1V.
Risk of hearing damage
To protect your hearing, make sure that the volume setting is not too high before put-
ting on the headphones.
3.2.1.5
USB
The front panel provides three female USB connectors (USB-A) to connect devices like
a keyboard (recommended: R&S PSL-Z2, order number 1157.6870.03) or a mouse
(recommended: R&S PSL-Z10, order number 1157.7060.03). In addition, a memory
stick can be connected to store and reload instrument settings and measurement data.
The rear panel provides further USB connectors, including a male (USB-B) connector.
See
All USB connectors support standard 2.0.
3.2.1.6
POWER SENSOR
The LEMOSA female connector is used to connect power sensors of the R&S NRP-
Zxy family. For a detailed list of supported sensors see the data sheet.
For details on configuring and using power sensors, see
3.2.1.7
PROBE
The R&S
FSW provides a connector for supply voltages of +15 V to -12 V and ground
for active probes and preamplifiers. A maximum current of 140 mA is available. This
connector is suitable as power supply for high-impedance probes.
3.2.1.8
NOISE SOURCE CONTROL
The noise source control female connector is used to provide the supply voltage for an
external noise source, e.g., to measure the noise figure and gain of amplifiers and fre-
quency converting devices.
Conventional noise sources require a voltage of +28 V in order to be switched on and
0 V to be switched off. The output supports a maximum load of 100 mA.
Instrument Tour