![Rohde & Schwarz FSW13 User Manual Download Page 785](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847785.webp)
Remote Commands
R&S
®
FSW
785
User Manual 1173.9411.02 ─ 19
Remote commands exclusive to statistical results
CALCulate<n>:STATistics:CCDF:X<t>?
........................................................................... 785
CALCulate:STATistics:RESult<t>?
...................................................................................785
CALCulate<n>:STATistics:CCDF:X<t>?
<Probability>
This command queries the results of the CCDF.
Query parameters:
<Probability>
P0_01
Level value for 0.01 % probability
P0_1
Level value for 0.1 % probability
P1
P1: Level value for 1 % probability
P10
Level value for 10 % probability
Example:
CALC:STAT:CCDF:X1? P10
Returns the level values that are over 10 % above the mean
value.
Usage:
Query only
Manual operation:
See
CALCulate:STATistics:RESult<t>?
<ResultType>
This command queries the results of a CCDF or ADP measurement for a specific
trace.
Parameters:
<ResultType>
MEAN
Average (=RMS) power in dBm measured during the measure-
ment time.
PEAK
Peak power in dBm measured during the measurement time.
CFACtor
Determined crest factor (= ratio of peak power to average
power) in dB.
ALL
Results of all three measurements mentioned before, separated
by commas: <mean power>,<peak power>,<crest factor>
Example:
CALC:STAT:RES2? ALL
Reads out the three measurement results of trace 2. Example of
answer string: 5.56,19.25,13.69 i.e. mean power: 5.56 dBm,
peak power 19.25 dBm, crest factor 13.69 dB
Usage:
Query only