Remote Commands
R&S
®
FSW
773
User Manual 1173.9411.02 ─ 19
Parameters:
<SweepTime>
Sweep time.
The range depends on the ratios of the span to the RBW and
the RBW to the VBW. Refer to the data sheet for more informa-
tion.
Manual operation:
See
[SENSe:]LIST:RANGe<range>:SWEep:TIME:AUTO
<State>
This command turns automatic selection of the sweep time for a spurious emission
measurement range on and off.
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Manual operation:
See
[SENSe:]LIST:RANGe<range>:TRANsducer
<Transducer>
This command selects a transducer factor for a spurious emission measurement
range.
Note that
●
the transducer must cover at least the span of the range
●
the x-axis has to be linear
●
the unit has to be dB
Suffix:
<range>
.
1..30
Selects the measurement range.
Parameters:
<Transducer>
String containing the transducer file name, including the path
information.
Manual operation:
See
11.5.7.3
Configuring the List Evaluation
The following commands configure the list evaluation.
Useful commands for spurious emission measurements described elsewhere
●
Configuring and Performing Measurements