Remote Commands
R&S
®
FSW
678
User Manual 1173.9411.02 ─ 19
*CAL?
Calibration query
Initiates a calibration of the instrument and subsequently queries the calibration status.
Responses > 0 indicate errors.
Usage:
Query only
Manual operation:
See
*CLS
Clear status
Sets the status byte (STB), the standard event register (ESR) and the
EVENt
part of
the
QUEStionable
and the
OPERation
registers to zero. The command does not
alter the mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
*ESE
<Value>
Event status enable
Sets the event status enable register to the specified value. The query returns the con-
tents of the event status enable register in decimal form.
Parameters:
<Value>
Range:
0 to 255
*ESR?
Event status read
Common Commands