Remote Commands
R&S
®
FSW
689
User Manual 1173.9411.02 ─ 19
Parameters:
<Mode>
SINGle
Each measurement is performed once (regardless of the chan-
nel's sweep mode), considering each channels' sweep count,
until all measurements in all active channels have been per-
formed.
CONTinuous
The measurements in each active channel are performed one
after the other, repeatedly (regardless of the channel's sweep
mode), in the same order, until the Sequencer is stopped.
CDEFined
First, a single sequence is performed. Then, only those channels
in continuous sweep mode (
INIT:CONT ON
) are repeated.
*RST:
CONTinuous
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Manual operation:
See
SYSTem:SEQuencer
<State>
This command turns the Sequencer on and off. The Sequencer must be active before
any other Sequencer commands (
INIT:SEQ...
) are executed, otherwise an error will
occur.
A detailed programming example is provided in
chapter 11.4.3, "Programming Exam-
ple: Performing a Sequence of Measurements"
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S
FSW User Manual.
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands (
INIT:
SEQ...
) are not available.
*RST:
0
Selecting the Operating Mode and Application