Remote Commands
R&S
®
FSW
1003
User Manual 1173.9411.02 ─ 19
Example:
INST:SEL 'SPECTRUM2'
Selects measurement channel
'SPECTRUM2'
.
MMEM:STOR:TYP CHAN
Specifies that channel-specific data is to be stored.
MMEM:STOR:STAT 1, 'C:\R_S\Instr\user\Spectrum'
Stores the settings from channel
'SPECTRUM2'
to the file
'C:\R_S\Instr\user\Spectrum'
.
MMEM:LOAD:TYPE NEW
Specifies that channel-specific settings are to be activated in a
new channel.
MMEM:LOAD:STAT 1, 'C:\R_S\Instr\user\Spectrum'
Loads the channel-specific settings from the file
'C:\R_S\Instr\user\Spectrum'
to the new channel
'SPECTRUM2*'
.
MMEMory:STORe:STATe
1,<FileName>
This command saves the current instrument configuration in a *.dfl file.
Secure User Mode
In secure user mode, settings that are to be stored on the instrument are stored to vol-
atile memory, which is restricted to 256
MB. Thus, a "Memory full" error may occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details see
chapter 3.1.7, "Protecting Data Using the Secure User Mode"
For details see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Parameters:
1
<FileName>
String containing the path and name of the target file.
The file extension is .dfl.
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current device settings in the file
Save.dfl
.
Usage:
Event
SCPI confirmed
Manual operation:
See
See
MMEMory:STORe:STATe:NEXT
This command saves the current instrument configuration in a *.dfl file.
Managing Settings and Results