Remote Commands
R&S
®
FSW
875
User Manual 1173.9411.02 ─ 19
[SENSe:]CORRection:CVL:BAND
<Type>
This command defines the waveguide band for which the conversion loss table is to be
used. This setting is checked against the current mixer setting before the table can be
assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Band>
K | A | KA | Q | U | V | E | W | F | D | G | Y | J | USER
Standard waveguide band or user-defined band.
Note:
The band formerly referred to as "A" is now named "KA";
the input parameter "A" is still available and refers to the same
band as "KA".
For a definition of the frequency range for the pre-defined bands,
see
*RST:
F (90 GHz - 140 GHz)
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:BAND KA
Sets the band to KA (26.5 GHz - 40 GHz).
Manual operation:
See
[SENSe:]CORRection:CVL:BIAS
<BiasSetting>
This command defines the bias setting to be used with the conversion loss table.
Before this command can be performed, the conversion loss table must be selected
(see
This command is only available with option B21 (External Mixer) installed.
Parameters:
<BiasSetting>
numeric value
*RST:
0.0 A
Default unit: A
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:BIAS 3A
Manual operation:
See
See
[SENSe:]CORRection:CVL:CATAlog?
This command queries all available conversion loss tables saved in the
C:\r_s\instr\user\cvl\
directory on the instrument.
Setting Basic Measurement Parameters