Remote Commands
R&S
®
FSW
753
User Manual 1173.9411.02 ─ 19
Parameters:
<State>
ABSolute
Evaluates only limit lines with absolute power values
RELative
Evaluates only limit lines with relative power values
AND
Evaluates limit lines with relative and absolute power values. A
negative result is returned if both limits fail.
OR
Evaluates limit lines with relative and absolute power values. A
negative result is returned if at least one limit failed.
*RST:
REL
Example:
CALC:LIM:ESP:PCL:LIM ABS
Manual operation:
See
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>:MAXimum
<Level>
This command defines the upper limit of a particular power class.
Note:
●
The last power class always has an upper limit of 200
dBm.
●
The upper limit of a power class must always be the same as the lower limit of the
subsequent power class.
●
The power class must already exist (see
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<class>
1...4
power class
Parameters:
<Level>
Range:
-199.9 dBm to 200 dBm
Example:
CALC:LIM:ESP:PCL1:MAX -40 dBm
Sets the maximum power value of the first power class to -40
dBm.
Manual operation:
See
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>:MINimum
<Level>
This command defines the lower limit of a particular power class.
Note:
●
The first power class always has a lower limit of -200
dBm.
Configuring and Performing Measurements