Using Gate Ranges for Statistical Measurements
The following commands control gated statistical measurements.
[SENSe:]SWEep:EGATe:TRACe<k>:COMMent
................................................................780
[SENSe:]SWEep:EGATe:TRACe:PERiod
......................................................................... 780
[SENSe:]SWEep:EGATe:TRACe<k>:STARt<range>
......................................................... 780
[SENSe:]SWEep:EGATe:TRACe<k>[:STATe<range>]
.......................................................781
[SENSe:]SWEep:EGATe:TRACe<k>:STOP<range>
..........................................................781
[SENSe:]SWEep:EGATe:TRACe<k>:COMMent
<Comment>
This command defines a comment for the gate of a particular trace.
Parameters:
<Comment>
String containing the comment.
Example:
SWE:EGAT:TRAC1:COMM 'MyComment'
Defines a comment for the gate in trace 1.
Manual operation:
See
[SENSe:]SWEep:EGATe:TRACe:PERiod
<Length>
This command defines the length of the gate for all traces.
The gate length applies to all traces.
Parameters:
<Length>
Range:
100 ns to 1000 s
*RST:
2 ms
Example:
SWE:EGAT:TRAC1:PER 5ms
Defines the period for gated triggering to 5 ms.
Manual operation:
See
[SENSe:]SWEep:EGATe:TRACe<k>:STARt<range>
<Time>
This command defines the start time for a gate range.
Suffix:
<range>
.
1...3
gate range