Remote Commands
R&S
®
FSW
784
User Manual 1173.9411.02 ─ 19
CALCulate<n>:STATistics:SCALe:Y:LOWer
<Magnitude>
This command defines the lower vertical limit of the diagram.
Parameters:
<Magnitude>
The number is a statistical value and therefore dimensionless.
Range:
1E-9 to 0.1
*RST:
1E-6
Example:
CALC:STAT:SCAL:Y:LOW 0.001
Manual operation:
See
CALCulate<n>:STATistics:SCALe:Y:UNIT
<Unit>
This command selects the unit of the y-axis.
Parameters:
<Unit>
PCT | ABS
*RST:
ABS
Example:
CALC:STAT:SCAL:Y:UNIT PCT
Sets the percentage scale.
Manual operation:
See
CALCulate<n>:STATistics:SCALe:Y:UPPer
<Magnitude>
This command defines the upper vertical limit of the diagram.
Parameters:
<Magnitude>
The number is a statistical value and therefore dimensionless.
Range:
1E-5 to 1.0
*RST:
1.0
Example:
CALC:STAT:SCAL:Y:UPP 0.01
Manual operation:
See
11.5.8.5
Performing a Statistical Measurement
The following commands are required to perform a statistical measurement:
chapter 11.5.1, "Performing Measure-
11.5.8.6
Retrieving Results
The following commands are required to retrieve the measurement results.
Useful commands for retrieving results described elsewhere:
●
Configuring and Performing Measurements