6
Choose the Freeze button from the Adjust Iris dialog box,
then return to step 4.
7
Choose the OK button to accept the calibration value,
then wait for “Calibration Completed” to appear. (The
system requires several seconds to calibrate itself for
color, brightness and contrast. The electronic flash fires
several times during the process.)
8
Choose the OK button to return to the Calibration dialog
box.
9
To calibrate the live image, go to
Calibrating the live
image on page 6-8.
To discontinue calibration and return to the System
Management window, choose the OK button and choose
Yes when the message “This will replace current
calibration. Continue?” appears. Then remove the
calibration target from the backdrop and store it where it
will not be marred or damaged.
6-10
ID-4000 input devices