Item
Function No.
Setting value Contents
0*
Invalid
Power on zero operation
CF-15
1
Valid
0*
Operation when stable and 0 < gross
weight ≤ maximum load
1
Unconditional operation and 0 < gross
weight ≤ maximum load
2
Operation when stable and 0 < gross
weight ≤ |maximum load|
Tare weight cancellation opera-
ting condition
CF-16
3
Unconditional operation and 0 < gross
weight ≤ |maximum load|
0*
RAM
Change of data storage destina-
tion
CF-17
1
EEPROM
Power on clear
CF-18
000 ~ 111
000*
0: Invalid
1: Valid
101: Tare weight cancellation data
102: Preset tare weight data
103: Zero set and zero tracking data
0*
Invalid
Stability detection during set
mode 2 or preset tare weight set
CF-20
1
Efective
0*
Invalid
Unit display in sub-display sec-
tion during set mode 2 or preset
tare weight set
CF-21
1
Efective
0*
Set the district number
Setting method for gravity acce-
leration compensation
CF-25
1
Set the numerical value for gravity accele-
ration
01 ~ 16
Unit: District
District number for place of use CF-26
(Possible to set
CF-25 = 0)
10*
01 ~ 16
Unit: District
District number for place of cali-
bration
CF-27
(Possible to set
CF-25 = 0)
10*
Gravity acceleration setting for
place of use
CF-28
(Possible to set
CF-25 = 1)
9.000 ~ 9.999
9.797*
Unit: m/s
2
Gravity acceleration setting for
place of calibration
CF-29
(Possible to set
CF-25 = 1)
9.000 ~ 9.999
9.797*
Unit: m/s
2
0*
Range change OFF
1
Changeable between two ranges
Range switch operation
CF-40
2
Changeable between three ranges
7 C function mode
Weight indicator CSD-903
Minebea Intec
EN-53
Summary of Contents for CSD-903
Page 243: ......