Note:
-
If an RS-232C interface is installed, function F-41 is valid if F-40 is set to [1 to 4].
-
If an RS422/485 interface is installed, function F-41 is valid if F-49 is set to [1].
18.3.5
Communication speciications by RS-232C and RS-422/485
RS-232C and RS-422/485 communication speciications are selected by function F-42.
The default is [13020]
Function No.
Set value
Details
F-42
00000 to 15121
0
0 0
0 0
0 : 7 bit, 1 : 8 bit
0 : 1 bit, 1 : 2 bit
0 : 1 200 bps,
1 : 2,400 bps
2 : 4 800 bps,
3 : 9 600 bps
4 : 19 200 bps
5 : 38 400 bps
0 : CR, 1 : CR + LF
Data bit length
Parity bit
Stop bit
Baud rate
Terminator
0 : No parity
1 : Even parity
2 : Odd parity
18.3.6
Address settings for the RS-422/485 interface
The address of the RS-422/485 interface is set by function F-43.
Setting range: 00 to 31
: 00
18.3.7
Changeover of RS-422/485
The changeover of RS-422 and RS-485 is executed by the setting of function F-44.
Setting range : 0, 1
0 : RS-422
1 : RS-485
18.3.8
Delay time for RS-485 return data
Return data can be delayed at the CSD-903side when the host-side transmission terminal
has low impedance after host-side communication is complete.
Delay time for return data from the RS-485 interface is set by function F-45.
Setting range: 0 to 9
: 01
Unit: 0.1 s
18.3.9
Transmission data decimal point usage for RS-232C and RS-422/485
Transmission data decimal point usage for RS-232C is selected by function F-46.
The default is [0].