10 Various operations by function data
10.1 Digital ilter
The digital ilter function stabilizes A/D converted data by moving average processing.
The moving average rate is selected by function F-01.
Setting range: 001 to 256 (Unit: Moving average frequency of 1 time)
The default is [16 times].
Characteristic trends of the running average are shown in the table below.
Running average frequency: Low/High
Running average frequency
Low
High
Noise resistance
Sharp
Stable
Response speed
Fast
Slow
10.2 Analog ilter
The analog ilter function cancels input signal noise.
The frequency is selected by function F-02.
The frequency can be selected from [2 Hz], [4 Hz], [6 Hz], [8 Hz] and [10 Hz].
The default is [4 Hz].
10.3 Display frequency
The display frequency is selected by function F-04.
The display frequency can be selected from [4 times/s] and [20 times/s].
The default is [4 times/s].
10.4 Stabilization ilter
The stabilization ilter function increases the digital ilter when the load display change is
ixed for longer than a prescribed time.
10.4.1
Stabilization ilter
The stabilization ilter moving average frequency is selected by function F-05.
Setting range: 001 to 256; Unit: Moving average frequency of 1 time
The default is [92 times].
10.4.2 Stabilization ilter data width
The stability detection data width is set by function F-06.
Setting range: 000 to 999; Unit: 0.1 D; 000: OFF
The default is [020].
10.4.3 Stabilization ilter time range
The stability detection data range is set by function F-07.
The stabilization ilter data width is set by function F-07.
Setting range: 00 to 99; Unit: 0.1 s; 00: OFF
Weight indicator CSD-903
10 Various operations by function data
EN-76
Minebea Intec