Number
Details
20
If the load cell output and numerical input is equal to or less than
–2.5 mV/V and exceeds the zero calibration range on the negative side
during calibration
21
If the load cell output and numerical input is equal to or more than
2.5 mV/V and exceeds the zero calibration range on the positive side
during calibration
22
If the load cell output and numerical input is less than 0 mV/V and does
not reach the span calibration range during calibration
23
If the load cell output and numerical input is more than 3.1 mV/V and
exceeds the span calibration range during calibration
31
EEPROM writing error
Hold registration (word read/write reference No. 4)
Address
Data name
1 to 9
UNUSED
11 to 12
FINAL/S1
13 to 14
OVERSHOOT/S4
15 to 16
PRIOR TO FIXED VALUE 1/S3
17 to 18
PRIOR TO FIXED VALUE 2/S2
19 to 20
OVER
21 to 22
UNDER
23 to 24
NEAR ZERO
25 to 26
FULL
27 to 28
PRESET TARE WIGHT
29 to 30
POST-BATCHING TIME
31 to 32
CALMING TIME AFTER POST-BATCHING
33 to 34
BRAND NO. 1 AUTOMATIC OVERSHOOT COMPENSATION
35 to 265
UNUSED
267 to 268
BRAND NO. 1 FINAL/S1
269 to 270
BRAND NO. 1 OVERSHOOT/S4
271 to 272
BRAND NO. 1 PRIOR TO FIXED VALUE 1/S3
273 to 274
BRAND NO. 1 PRIOR TO FIXED VALUE 2/S2
275 to 276
BRAND NO. 1 OVER
277 to 278
BRAND NO. 1 UNDER
279 to 280
BRAND NO. 1 NEAR ZERO
281 to 282
BRAND NO. 1 FULL
283 to 284
UNUSED