Set the load cell output voltage:
mV/V
[END]
▼ End.
Error display in span calibration
[SP-L]: The display blinks for about 2 seconds if ([Load cell output voltage
at span point] – [Load cell output voltage at zero point]) ≤ 0.0 mV/V (refer to
Chapter
[SP-H]: The display blinks for about two seconds if load cell output is 3.1
mV/V or more (refer to Chapter
).
6.3.8
Quit calibration
[END]
[END] is displayed when SPAN calibration is complete.
Stand-by condition
Note:
Set ine calibration for ZERO and SPAN, digital linearization, ZERO set and tare weight
cancellation after completing calibration. Recalibration clears set data.
6.4 Fine calibration of zero and SPAN
This function executes ine calibration of zero and SPAN when there is a diference
between the actual weight value and the weight.
6.4.1
Switch to zero and SPAN ine calibration mode
Weighing mode
Stand-by condition
[CFUNC]
3. Press the [F] key after setting
1. Press the [F] key to quit calibration mode.
[CADJ]
The display changes to [CADJ.], and the set data is saved to the internal memory.
2. Press the [ON-OFF] key to move to stand-by condition.
1. Press the [ON-OFF] key to move from normal weighing mode to the stand-by
condition.
6 Calibration
Weight indicator CSD-903
Minebea Intec
EN-43