①
Change to Zero point and SPAN point ine calibration mode
②
Zero stability check
③
SPAN point ine calibration
④
SPAN stability check
⑤
SPAN point ine calibration
⑥
Finish Zero point and SPAN point ine calibration
Note:
-
Calibrating by transmission
(-RS-) is displayed.
-
When the calibration is force inished, all data returns to its condition prior to
calibration without registration.
-
Be sure to register the data for calibration with a weight after conirming the
condition is stable.
17.1.2.1
Change to Zero point and SPAN point ine calibration mode
Shift to Zero point and SPAN point ine calibration mode by writing [10] in holding
register address 56833.
17.1.2.2
Zero stability check
The weighing values are read from holding register addresses 56857 to 56858.
Stable/Unstable is read from holding register address 56845 with no load on the load
cell.
0: Unstable
1: Stable
17.1.2.3
Zero point ine calibration
Zero the load values from holding register addresses 56857 to 56858.
Execute ine calibration after writing the data shown in below table in holding register
address 56849.
Holding register address 56849entry
Fine calibration on the positive side
1
Coarse calibration on the positive side
10
Fine calibration the negative side
–1
17 ModBus communication