10.5.2 Key lock 2
Function F-09 sets whether key operation of [TARE WEIGHT/▲], [NET/GROSS
WEIGHT/▼], [ZERO] or [F/
↵
] are[valid] or [invalid].
Setting range: 0000 to 1111; 0: Invalid; 1: Valid
10
0
digit: TARE WEIGHT/▲
10
1
digit: NET/GROSS WEIGHT/▼
10
2
digit: ZERO
10
3
digit: F/
↵
10.6 Stability detection
When the change width of the load display is within the data width set by function F-10,
and when the condition continues for the time set by function F-11 or more, the stability
detection is the function to judge the weighing value being stable.
Note:
When two range or three range switches have been selected by the automatic range
switches, the time width of the stability detection execute the display conversion of the
time width by the value of the irst range scale interval (scale interval in the calibration).
10.6.1
Stability detection data width
The stability detection data width is set by function F-10.
Setting range: 000 to 999; Unit: 0.1 D; 0: OFF
The default is [020].
The display conversion data width is calculated as shown below.
Stability detection data width = Set value of F-10 × 0.1 × scale interval
The F-10 setting is [5] and the scale interval is [D = 2].
Data width stability detection = 5 × 0.1 × 2 = 1 D
10.6.2 Duration of stability detection
The stability detection time width is deined by function F-11.
Setting range: 00 to 99; Unit: 0.1 s; 0: OFF
The default is [10].
Note:
When two range or three range switches have been selected automatically, the stability
detection data width executes data width display conversion by the irst range scale
interval (calibration scale interval).
Weight indicator CSD-903
10 Various operations by function data