In this mode, each output is turned ON, according to the value set by the ON limit of
the start signal (OFF → ON). If the set value is compared with the measurement value,
and the comparison result is satisied, then the output goes OFF.
In this mode, when an output turns OFF, it remains OFF until the inish signal is
output, even if the comparison result is outside the relevant condition
For more information refer to Chapter
This mode enables sequential control without connecting an external sequencer, etc.
- 4-steps check mode 1
Compares a set value with the weighed value and outputs the comparative result.
With 4-steps check mode 1, output turns ON or OFF depending on the [Under] or
[Over] status after comparison of set value 1 with the weighed value.
For more information refer to Chapter
.
- 4-steps check mode 2
With 4-steps check mode 2, weighed values can be sorted because the output turns
on when a weighed value is within two set points.
For more information refer to Chapter
.
11.1 Weighing mode setting method
Weighing mode
[FUNC.]
[SQFUNC]
[SQF-01]
Select the SQ function number to be changed.
[TARE WEIGHT] [NET-GROSS WEIGHT]: Adjust the value of the digit to be changed.
[ACCUM] [PRESET TARE WEIGHT]: The changed digit is selected.
[ZERO]: The displayed value changes to [0].
[SET]: The display returns to [SQFUNC].
[F]: The displayed value is saved. Proceed to the next step.
[ON-OFF]: Quit SQ function mode without changing other settings.
1. Change from normal weighing mode to the [FUNC] display by the [SET] key.
2. Press the [NET-GROSS WEIGHT] key to display [SQFUNC] and switch to SQ function
mode.
3. Press the [F] key[ to display SQF-01].
4. Press the [F] key. Set the value of the selected [SQF-01].
11 Weighing mode
Weight indicator CSD-903
Minebea Intec
EN-89