District
No.
Acceleration
(m/s
2
)
Corresponding districts
7
9.800
Fukushima Prefecture, Ibaraki Prefecture, Niigata Prefecture
8
9.799
Tochigi Prefecture, Toyama Prefecture, Ishikawa Prefecture
9
9.798
Gunma Prefecture, Saitama Prefecture, Chiba Prefecture, Tokyo (excluding the
jurisdiction of the branch administrative oices Hachijojima and Ogasawara), Fukui
Prefecture, Kyoto Prefecture, Tottori Prefecture, Shimane Prefecture
10
9.797
Aichi Prefecture, Mie Prefecture, Wakayama Prefecture, Shiga Prefecture, Osaka
Prefecture, Hyogo Prefecture, Nara Prefecture, Okayama Prefecture, Hiroshima
Prefecture, Yamaguchi Prefecture, Tokushima Prefecture, Kagawa Prefecture
11
9.796
Tokyo (jurisdiction of branch administrative oice of Hachijojima only), Ehime Pre-
fecture, Kochi Prefecture, Fukuoka Prefecture, Saga
Prefecture, Nagasaki Prefecture, Oita Prefecture
12
9.795
Kumamoto Prefecture, Miyazaki Prefecture
13
9.794
Kagoshima Prefecture (excluding jurisdiction of branch administrative oices of
Nase and Oshima)
14
9.793
Tokyo (jurisdiction of branch administrative oice of Ogasawara only)
15
9.792
Kagoshima Prefecture (jurisdiction of branch administrative oice of Nase and
Oshima only)
16
9.791
Okinawa Prefecture
Note:
Conirm the details of gravitational acceleration using the service for gravity data on the
websites of "Science chronology," the Geographical Survey Institute, etc.
(http://vldb.gsi.go.jp/sokuchi/gravity/grv_search/gravity.pl)
8.14 Automatic range switch
The automatic range switch function switches automatically and displays the scale
interval by the weighing data of the gross or net value by dividing the range of weighing
to the maximum load into 2 or 3.
8.14.1
Setting the range switch operation
Range switch operation is selected by C function CF-40.
The range switch can be selected from [OFF (without range switch)], [Two range switch]
and [Three range switch].
The default is [OFF].
8.14.2 Setting the scale interval of the second range
The scale interval of the second range is set by C function CF-41.
The scale interval can be selected from [1 scale interval], [2 scale interval], [5 scale
interval], [10 scale interval], [20 scale interval] and [50 scale interval].
The default is [5 scale interval].
Weight indicator CSD-903
8 Various operations by C function data
EN-62
Minebea Intec