Excitation
System coniguration
10 V
5 V
2.5 V
5 load cells + cable junction box + Zener barrier
×
×
◯
6 load cells + cable junction box + Zener barrier
×
×
×
◯
Applicable; × Not applicable
The above examples use 6-wire cable and 350
Ω
load cells.
8.6 Net weight sign reversal function
The net weight sign inversion function inverts the sign of the net weight display and
external net weight data output, for example, in simple discharge mode.
The net weight sign inversion function makes the net weight display or external net
weight data output positive by inverting the sign of the net weight using C function
CF-08.
Net weight sign inversion can be selected from [Net weight sign not inverted] or [Net
weight sign inverted].
The default is [Net weight sign not inverted].
8.7 Zero set
The ZERO set function stores the present measuring value as the zero point and calibrates
the display to zero by pressing the [ZERO] key when the value indicated in the load
display is the gross value and is within the zero set efective range.
8.7.1
Operating condition of ZERO set
The zero set operating condition is selected by C function CF-10.
The efective range is shared by both zero set and zero tracking.
[Operation under stable conditions] or [Unconditional operation] can be selected.
The default is [Operation under stable conditions].
8.7.2
Zero set efective range
The Zero set efective range is set by C function CF-11.
Setting range: 00 – 30; Unit : ±1% of the weighing capacity (maximum load); 00: OFF
The default is [02].
Note:
-
When zero compensation is executed within its efective range, with zero set and
zero tracking, zero set is not accepted.
-
When tare weight cancellation or preset tare weight cancellation is executed, zero
set is executed after clearing the tare weight cancellation and the preset tare weight
cancellation.
8 Various operations by C function data
Weight indicator CSD-903
Minebea Intec
EN-57