11.3.2
Simple discharge mode
Simple discharge mode outputs a judgment condition that is satisied in comparison
between the decrease in weighing amount for discharge of the weighed object and a set
amount.
Judgment condition
Status display
Judgment condition
Z-BAND
|Gross weight or Net weight)| ≤ ZERO BAND (SET7)
F. (Coarse) FLOW
Net weight ≥ [Final (SET1) – Prelim 2 (SET4)]
M. (Middle) FLOW
Net weight ≥ [Final (SET1) – Prelim 1 (SET3)]
D. (Fine) FLOW
Net weight ≥ [Final (SET1) – Prelim 1 (SET3)]
UNDER
Net weight < [Final (SET1) – Under (SET6)]
OVER
Net weight > [Final (SET1) + Over (SET5)]
OK
[Final (SET1) – Under (SET6) ≤ Net weight ≤ Final (SET1) + Over
(SET5)]
Operating explanations
-
When –net weight reaches [Final value – Prelim 2] Coarse low turns ON. When the
start signal is input, the comparison from Coarse low, Middle low and Fine low is
prohibited during the Middle low comparison.
-
When the –net weight reaches[Final value – Prelim 1] Middle low turns ON. When the
start signal is input, the comparison from Coarse low, Middle low and Fine low is
prohibited during the Fine low comparison.
-
When the net weight reaches [Final – Freefall], Fine low turns ON.
-
When the start signal is used, the inish signal is output.
-
The comparative target of near zero and the comparative target of FULL value
can be selected from Gross weight and Net weight.
-
Under, Over and OK operate with net weight.
Comparison operation can be changed to simple operation.
-
When [Prelim 2 = 0]is set, Coarse low turns OFF; when [Prelim 1 = 0], Middle low
turns OFF.
Note:
-
If the start signal is not used, each condition display turns on and of according to
the judgment condition shown above.
-
Output for each comparison is synchronized with A/D sampling.