Under normal operation
Return (device → host)
10
8
6
4
2
12
14
@
0
6
0
0
Terminator
Command No.
Address
00 ~ 31
Set value
Sign
Under abnormal operation
Return (device → host)
6
6
8
4
2
@
0
0
E
R
1
3
Terminator
Address
00 ~ 31
Note:
-
Values are entered from the right.
-
Do not add a decimal point.
-
Set the storage destination to Internal RAM when the set value is continuously
changed and the number of EEPROM rewrites might exceed 1 000 000.
-
Do not change in the setting during weighing because the A/D sampling rate
decreases temporarily while rewriting the set value. Set the storage destination to
Internal RAM to facilitate changes setting values during weighing.
-
To write the efective brand number for [Command 69], set function F-51 to [2].
-
Weighing setting data writing allows the setting to be changed for the brand
number selected by [Command 6A].
Note:
The error transmission condition is as follows:
-
There is a disagreement in scale interval.
-
A value that exceeds the regular value is set
-
A non-numerical character is set.
-
A character other than + or - is set under Sign.
18.3.12.10 Writing comparison set data (host → device)
Command No.
Operation
60
Writing S1 setting
61
Writing S4 setting
18 Options
Weight indicator CSD-903
Minebea Intec
EN-189
Summary of Contents for CSD-903
Page 243: ......