Note:
-
While executing the calibration by transmission,
(-RS) is displayed.
-
When the calibration is compulsorily inished, all the data returns to the condition
before the calibration without executing the registration.
-
Please register the data for the calibration with the weight after conirming the
stable condition without fail.
17.1.1.1
Change to calibration mode by communication
Input [9] at holding register address 56833 to switch to calibration mode.
17.1.1.2
Setting the scale interval
The scale interval is written in holding register address 56835 to 56836.
Scale interval: Can be selected from 1, 2, 5, 10, 20 and 50
Error outside the setting range
When data outside the setting range is entered, the following codes are set.
Condition
Error code
Contents
Illegal data
0x03
When data outside the setting range is entered
17.1.1.3
Set the maximum load
The maximum load is written in holding register address 56837 to 56838.
Setting range: 1 to 99999
Error outside the setting range
When data outside the setting range is entered, the following codes are set.
Condition
Error code
Contents
Illegal data
0x03
When data outside the setting range is entered
17.1.1.4
Set the weight mass
Mass of weight is written in holding register address 56839 to 56840.
Setting range: 1 to 99999
However, data is below the maximum load.
Error outside the setting range
When data outside the setting range is entered, the following codes are set.
Condition
Error code
Contents
Illegal data
0x03
When data outside the setting range is entered
17.1.1.5
Conirm zero stability (actual weight)
Stable/Unstable is read from holding register address 56845.
0: Unstable
1: Stable