22: Force-inishes weighing
operation
Inputs the signal to force-inish weighing operation
23: Force-inish low
Inputs the signal to force-inish low
24: Accumulation clear
Clears the accumulation value and the accumulation
times of all brands
25: Brand 1
Inputs brand number using BCD code
26: Brand 2
Inputs brand number using BCD code
27: Brand 4
Inputs brand number using BCD code
The default is [OFF].
The following level input items are enabled with a bridge input of 50 ms or more.
-
[13: Inlow and discharge changeover]
-
[18: Hold]
-
[21: Net weight display]
Other items are efective a pulse input of 50 ms or more.
10.14.2 External control output setting method
The external control output operation is selected by functions F-70 to F82.
Setting range: 00 to 22
0: OFF
No setting
1: STABLE
Output when stable
2: DURING TARE WEIGHT
Output during tare weight cancellation
3: GROSS WEIGHT DISPLAY
Output when gross weight is displayed
4: NET WEIGHT DISPLAY
Output when net weight is displayed
5: CENTER OF ZERO
Output in the zero point state
6: DURING HOLD
Output when hold is applied
7: ERROR
Output when an error occurs
8: ABNORMAL WEIGHT
Output when display is OL
9: WEIGHING Sequence
Output when a sequence error occurs
10: NEAR ZERO
Output in near zero state.
11: OK/S0
Lights when the judgment condition is OK in sequential
mode Lights when S0 output signal is ON in check mode
1 or 2
12: F. (Coarse) FLOW/S1
Lights when coarse low output signal is [ON] in sequen-
tial mode Lights when S1 output signal is[ON]in check
mode 1 or 2
13: M. (Middle) FLOW/S2
Lights when middle low output signal is [ON]in sequen-
tial mode Lights when S2 output signal is [ON]in check
mode 1 or 2
Weight indicator CSD-903
10 Various operations by function data
EN-86
Minebea Intec