6.4.1.2
SPAN ine calibration
[SPAN]
Place a known weight (calibration weight) of less than the maximum load on the load
cell.
[
✶✶✶✶✶
.
✶
] is the current weight value. It can display up to 1/10 digit.
Set the display to the same value with the weight on the load cell.
[TARE WEIGHT]: Increase the value Increase continuously by holding down the key.
[NET-GROSS]: Decrease the value Decrease continuously by holding down the key.
Proceed to "Exit ine calibration mode" after SPAN ine calibration is complete.
[END]
▼ End.
6.4.1.3
Completing ine calibration mode
[END]
Stand-by condition
Zero and SPAN ine calibration is complete.
[SPAN]
Precedes operation of "Span ine calibration" if executing span ine calibration.
Otherwise, press the [NET-GROSS] key, and proceed to "Exit ine calibration
mode" after [END]is displayed.
▼ Span ine calibration.
1. [
✶✶✶✶✶
.
✶
] is displayed by pressing the [F] key.
2. After [END]is displayed, press the [F] key.
1. Press the [F] key to quit zero and SPAN ine calibration mode.
[LNR]
The set data is saved to the internal memory.
[LNR]is displayed.
2. Press the [ON-OFF] key to move to the stand-by condition.