[P-1]
The display moves to [P-1] → [P-3] by pressing the [TARE WEIGHT] key. Select the
numbers of the points for linearization compensation, and press the [F] key.
[P-1]: Compensation by 1 point
[P-2]: Compensation by 2 points
[P-3]: Compensation by 3 points
[LNR-1]
Load the weight
Conirm that the stability mark light is on.
When the 100 digit of the weighing value blinks, set to the same value with the
weight on the load cell.
[TARE WEIGHT]: Increase the weighing value
[NET-GROSS]: Decrease the weighing value
[F]: The displayed value is memorized. Proceed to the next step.
[SET]: Interrupt the setting and return to the [LNR] display.
The display moves to[LNR-2] when [P-2] is selected. Execute compensation for the
second point.
The display moves to[LNR-3] when [P-3] is selected. Execute compensation for the
third point.
Ensure that the value of weights [LNR-1] < [LNR-2] < [LNR-3].
1. Pressing the [F] key twice from the[LNR] display to display[P-1].
2. After the display shows [LNR-1], load the weight of the point to be compensated on
the load cell and press the [F] key.
3. When the weighing value blinks, press the [F] key after conirming the stability mark
(▼) light is on.
4. Press the [F] key when compensation is complete for each point.
[END]
[END] is displayed.
5. Press the [F] key to complete digital linearization.
[TARE WEIGHT]
[TARE WEIGHT] is displayed. The set data is saved to the internal memory.
6 Calibration
Weight indicator CSD-903
Minebea Intec
EN-47